    LI Zhen-lei, LI Wei-hua. Working Performance of Chloride Ion Sensor Fabricated by Compression Method[J]. Corrosion & Protection, 2015, 36(6): 547-550. DOI: 10.11973/fsyfh-201506009
    Citation: LI Zhen-lei, LI Wei-hua. Working Performance of Chloride Ion Sensor Fabricated by Compression Method[J]. Corrosion & Protection, 2015, 36(6): 547-550. DOI: 10.11973/fsyfh-201506009

    Working Performance of Chloride Ion Sensor Fabricated by Compression Method

    • A chloride ion sensor with Ag/AgCl as working electrode and MnO2 as reference electrode was fabricated by a compression method.A series of tests were conducted to examine the performance of the sensor.The test results showed that the stability of the sensor was excellent.The Nernst equation test showed a strong linear correlation.Temperature had little effect on the results,but the presence of other ions had a certain impact on the sensor's potential value.The resistance of the sensor against the constant current polarization was weak.